Replacing The Spray Water Cartridge - Dentsply Sirona Midwest Stylus Plus Instrucciones De Uso

Ocultar thumbs Ver también para Midwest Stylus Plus:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6607290.book Seite 62 Donnerstag, 4. Januar 2018 11:55 11
8 Maintenance
A
B
62
Operating Instructions Midwest Stylus
➢ Remove the LED and re-insert it after rotating it 180° around its own
axis.
8.2.4

Replacing the spray water cartridge

Required accessories
● Spray water cartridge
● Midwest Plus Aerosol Spray
If the spray water cartridge leaks or is clogged.
1. Remove the quick coupling from the supply hose.
2. Remove the sealing washer (B).
3. Click the control ring into the maximum water flow. The spray water
C
cartridge (A) can be replaced only in this position.
4. Remove the spray water cartridge.
5. Spray oil lightly onto the new spray water cartridge.
6. Insert the new spray water cartridge.
®
®
Plus / Midwest Tradition
TC/Pro TCF
66 07 290 D3673.201.01.04.02

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Midwest tradition tcMidwest tradition pro tcf

Tabla de contenido