6607290.book Seite 52 Donnerstag, 4. Januar 2018 11:55 11
7 Conditioning
52
Operating Instructions Midwest Stylus
7.4.3
Care of cooling spray nozzle openings
If your tap water is very hard, lime deposits may constrict or completely
block the cooling spray nozzles.
Interval
Interval
● At least once a week
1. Carefully clean the nozzle openings by running a cleaning wire
through them.
2. Let the high-speed handpiece run briefly with cooling spray.
®
®
Plus / Midwest Tradition
TC/Pro TCF
66 07 290 D3673.201.01.04.02