Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Error Messages and Trouble Shooting
About early applying blood sample or control solution
The blood sample or control solution are applied before the meter is ready. Repeat the test with a new
test strip. Add sample after the "
" symbol appears.
About Strip Error - Er1
Test strip inserted has been used or damaged. Please use a new test strip from vial.
About Meter Error - Er2
Meter has malfunctioned. Do the quality control test or reinstall the batteries to see if the meter works
properly.
If error screen still appears, contact Bionime Customer Service.
About Signal Error - Er3
When you do the test, the meter find the signal is unreasonable. Please do a new test again.
60
- Store the strips in the original capped vial at temperatures between 4℃ to 30℃ ( 39℉ to 86℉ ) and relative humidity
below 90%. Do not freeze.
®
- Rightest
Blood Glucose Test strips are designed for using with capillary whole blood and venous whole blood
samples. Do not use serum or plasma samples.
- Inaccurate test results may be obtained at high altitude more than about 10000 feet ( 3048 meters ) above sea level.
- Hematocrit below 30% may cause higher results, and hematocrit above 55% may cause lower results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
®
- Rightest
Blood Glucose Monitoring System has not been validated for use on neonates.
- The glucose test may be interfered under abnormal concentration of
Uric acid > 9.0 mg/dL
L - Dopa > 1.5 mg/dL
Methyldopa > 1.5 mg/dL
Cholesterol > 250 mg/dL

Limitations

61

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para BIONIME GM300

Tabla de contenido