Use
Problem
The LED on the charging bay
quickly flashes red
The LED on the charging bay
flashes red, 3 times every 5
seconds
The LED on the charging bay
flashes red, 5 times every 5
seconds.
No LEDs are lit on the battery
charger
Batteries are not being charged

6.3 Information for use

INFORMATION
Lithium-ion batteries
The product is powered by a lithium-ion battery. Special requirements apply for this battery type.
► Travel: Before a trip, check the requirements of public authorities and transport companies (such as airlines).
For example, a battery on its own may have to be transported in carry-on luggage.
► Damaged batteries: Contact your O&P professional for information on transporting a damaged battery.
Checking the charge level
► Briefly press the button on the battery to activate the LEDs for the charge level.
LEDs on the battery
LED shows green light
LED is flashing red
4 angle symbols are on steady
3 angle symbols are on steady
2 angle symbols are on steady
1 angle symbol is on steady
1 angle symbol is flashing
Sleep mode
The battery of the prosthetic foot has a sleep mode. Sleep mode is activated to protect the battery, for example
when:
The temperature is too high (45 °C during charging, 65 °C during use)
The temperature is too low (less than 5 °C during charging, no limit during use)
The battery is drained (protection against deep discharge)
The prosthetic foot can still be used, but no longer has any active functions.
1) If battery sleep mode has been activated: Press and hold the button on the battery to restart the prosthetic
foot.
→ The LED on the battery flashes red: There is a problem with the battery. Contact your O&P professional.
→ The LEDs on the battery do not respond: Charge the battery.
2) If the temperature was too high: Let the battery cool down.
36
Solution
Take out the battery, disconnect the battery charger from the mains net­
work, connect the battery charger to the mains network again and reinsert
the battery. If the red flashing continues, contact your O&P professional.
Indicates a battery fault. Take the battery out of the battery charger and do
not use it again.
Contact your O&P professional.
Indicates that the temperature has been exceeded. Make sure the ventila­
tion slots of the battery charger are not blocked, and that the battery char­
ger is in a cool location. Disconnect the battery charger from the mains net­
work and allow it to cool before using it again.
The battery charger is not connected to the mains network
Check whether the power supply is connected to the battery charger and
the outlet is supplied with power.
Check whether the battery charger is connected to the mains network.
Check whether the battery is properly inserted and is at the correct operat­
ing temperature.
Try the other charging bay on the battery charger.
If the error could not be corrected, contact your O&P professional.
Status
The prosthetic foot is switched on.
There is an error. The prosthetic foot does not turn on.
Charge level: 76 % to 100 %
Charge level: 51 % to 75 %
Charge level: 26 % to 50 %
Charge level: 11 % to 25 %
Charge level: <10 %
Empower 1A1-1