► Ensure that the approved service life is not exceeded.
► Only use the product on a single patient.
CAUTION
Mechanical damage to the product
Risk of injury due to change in or loss of functionality
► Use caution when working with the product.
► If the product is damaged, check it for proper function and readi
ness for use.
► In case of changes in or loss of functionality, do not continue
using the product (see "Signs of changes in or loss of functional
ity during use" in this section).
► Take any necessary measures (e.g. repair, replacement, inspec
tion by the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Among other factors, changes in functionality can be indicated by an
altered gait pattern, a change in the positioning of the prosthetic com
ponents relative to each other and by the development of noises.
4 Scope of delivery
Quant
Designation
ity
1
Instructions for use
1
Lamination anchor
1
Shuttle lock housing
1
Ratchet unit
1
Pin
1
Plaster protection for pin
Quant
ity
1
Only for 6A20=20:
4
5 Preparation for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connec
tions
► Clean the threads before every installation.
► Apply the specified torque values.
► Follow the instructions regarding the length of the screws and
Reference
about how to secure the screws.
number
–
CAUTION
4R111=N
Improper processing of the anchor arms of the lamination
–
anchor
6A52
Risk of injury due to breakage of the anchor arms
6A20=10:
► Only bend anchor arms made of steel.
6Y13=1
► Avoid excessive or too frequent bending of the anchor arms.
6A20=20:
► Use the 711S4* bending irons for bending.
6Y13=2
5X440
Designation
Dummy set (1 hemispherical dummy, 1
screw dummy, 1 plaster screw)
Set screw
Reference
number
5X55
506G3
11