Mains voltage
Current consumption
Mains fuses
Patient leakage current
Ditto, first fault condition
Earth (ground) leakage current
Ditto, single fault condition
Medical device classification
Safety class
Weight
Dimensions
Safety tests
Environment conditions for transport and storage
Environment temperature
Relative humidity
Atmospheric pressure
Environments conditions for normal use
Environment temperature
Relative humidity
Atmospheric pressure
Classification
Medical class IIb
This equipment complies with all requirements of the Medical Device Directive (93/42/EEC).
International Safety class IEC 60601-1
*I
: The equipment has a safety earth (ground) connection, and must be connected to
an earthed (grounded) wall socket.
**BF : The equipment has a floating patient circuit. The Endomed 581
quirements of IEC 60601-1 and IEC 60601-2-10.
Technical modifications reserved.
For the ordering data of the Endomed 581
logue for Physiotherapy.
: 115 V , ± 15% 50-60 Hz
220-240 V , ± 10% 50-60 Hz
: 0.13 A (230 V )
: 2 x T1.6H250V
: typically 10 µA (requirement IEC eis < 100 µA)
: typically 20 µA (requirement IEC eis < 500 µA)
: typically 30 µA (requirement IEC eis < 500 µA)
: typically 60 µA (requirement IEC eis < 1000 µA)
: IIb (according to Medical Device Directive (93/42/EEC))
: I* type BF**, according to IEC 60601-1
: 5,5 kg
: 36 x 33,5 x 12 cm
: TÜV Rheinland
: -10° till +50° C
: 10 till 95 %
: 500 till 1060 hPa
: 10° till 40° C
: 10 till 90 %
: 500 till 1060 hPa
ID
, standard accessories and additional accessories we refer to the Cata-
TECHNICAL DATA
ID
meets the re-
Type BF
ORDERING DATA
27