AD2817C Operation Manual
2.5 Test Signal
2.2.1 Test singal frequency
10 typical frequencies: 50Hz, 60Hz, 100Hz, 120Hz, 1kHz, 10kHz, 20kHz, 40kHz, 50kHz,
100kHz.
Frequency accuracy: 0.02%
2.2.2 Test signal level
3 test levels: 0.1V, 0.3V, 1V
Accuracy:
(10% set value+2mV)
2.2.3 Output impedance
30
5%, 100 5% (fixed at 100 source impedance at bias mode)
2.2.4 Bias
Internal bias voltage: fixed +2V, -2V
Internal bias voltage: externally adjusted -5V +5v
If bias current mode is selected, the value of bias current is
Vb
Ib
Rs
Rx
Where, Vb is bias voltage, Rs is source resistance (100 ), Rx is DC resistance of the device
under test.
Bias current is adjustable within the range from -50mA to +50mA.
2.2.5 Measurement display range
Parameter
DCR, R2
L2A, L2B
If non-percentage data is out of display range, it will be displayed as 99999.
L
C
R, X, Z
D
Q
Angle: -179.99° 179.99°
θ
Radian: -3.1415 3.1415
%
Display range
0.001 H 99.999kH
0.001pF 9999.9 F
0.0001
99.999M
0.01m
1.9999M
0.001 H 99.999kH
0.0001 9.9999
0.0001 9999.9
-99.999% 999.99%
10