10.2.2
Sensor Head Inspections
Verify the absence of the following:
10.2.3
Cables Inspections
Verify the absence of the following:
10.3
Problems Found During Inspection
If inspections turn up serious problems, remove the sonar process
monitoring system from service until repairs can be made.
.
20909-01C Rev 06
• Loose or damaged latch hardware
• Loose or damaged boot seal clamps
• Damage to enclosure
• Damage to seals
• Loose or damaged access cover (including screws and seals)
• Loose handles
• Loose or damaged electrical connector or seals at the
connector
• Evidence from the condition of the pipe that a pipe leak under
the Sensor Head might occur before the next inspection
(reference paragraph 7.6).
• Damaged cables
• Improperly installed cables (for example, cable is under tension
or too tight bend radius)
• Loose cable management elements (strain reliefs, clamps, etc.)
Page 10-2