7.5 Operating test :

x
This test must be performed prior to each sterilisation.
connect the Medical Device to the motor ; point the head of the
Medical Device downwards. operate the Medical Device slowly
(5 000 rpm motor) for 30 seconds in order to get rid of the excess
oil, then at full speed (25 000 rpm motor) for 30 seconds. Wipe
the Medical Device with a disinfectant wipe if the oil has dripped.
noTE : if you notice a rise in temperature, irregularities,
vibrations, or abnormal noises during operation of the Medical
Device, contact your approved technical service immediately.
7.6 Sterilisation :
F
The Medical Device is supplied unlubricated and unsterilised.
The Medical Device should be sterilised before they are first
used and following each use.
• only sterilise instruments which have previously been disinfected,
cleaned, lubricated and tested.
• The Device should be sterilised at 135°c at 2.13 bars (275°f
at 30.88 psi), in a steam steriliser only, for a minimum duration
of 20 minutes (time for which sterilisation must be continued).
• We strongly recommend the use class B autoclave .
D
• Any other method of sterilisation should be avoided.
• Read the instruction leaflet provided by the steriliser
manufacturer.
• Remove the instrument from the Medical Device prior to
sterilisation.
• use sterilisation pouches suitable for Medical Device and the
steriliser, in accordance with the standard nf En 868. Always use one
single Medical Device per pouch.
• Adhere to the space specified between pouches and do not overload
the steriliser.
• Make sure that the device does not have any areas of corrosion or
cracks, and check that it is operating properly. Ensure that the product
is dry ; if necessary, dry any residual water with medical quality
pressurised air.
• After each sterilisation cycle, check that there is no water remaining
on the inside and outside of the packaging. Make sure that the flow
indicator has changed to the correct color.
• In order to avoid any retention of water, place the pouch in the
steriliser in such a way that any concave parts are face down.
• Keep the devices in the sterilisation pouches away from light,
moisture and any contamination. follow the manufacturer's
recommendations as seen on the packaging.
• The duration for which the device is kept after sterilisation should
not exceed 1 month. Label the devices, specifying the expiration date.
After the expiration date, repeat the cleaning and sterilisation
procedure.
In the event of breakdown, please contact your approved distributor or
the Anthogyr after sales sevice department directly.
E
All repairs must be carried out with parts and subassemblies
approved by the manufacturer.
VIII. REPAIR