8.5  Error limits
Message
Out of range, determine again
Buffer/standard temp. out of range
Offset out of range
Slope out of range
Wrong buffer
Invalid pH for user-defined buffer
ATC measured temperature is different
to the user-defined value
38
Maintenance
Description and Resolution
The software update process failed. This could be due to
the following reasons:
The USB stick is not connected or it is disconnected
during the update process
The update software is not in the correct folder
The exporting process failed. This could be due to the fol­
lowing reasons:
The USB stick is not connected or it is disconnected
during the exporting process
The USB stick is full
Range not accepted
pH
mV
T (pH)
| Eref1-Eb | > 60 mV
Slope < 85% or > 110%
I ∆Eref1I < 10 mV
I ∆pHI < 1 pH
| T(ATC)T(buffer) | > 1 °C
<-2.000 or > 20.000
<-2000.0 or > 2000.0
< 5 or > 50 °C