Setting Up the Monitor
NIBP Hose Connector
Biomed Ground Lug
AC Plug Receptacle
1
Connect only EN 60601-1 compliant devices, as specified by Philips.
2
The Biomed ground lug is an additional ground point that can be used by a Biomed when external grounding is
needed.
3
The handle may feel warm because it is also used as an exhaust port for the unit.
Step
Action
1
Connect the monitor to a grounded, 3-wire, hospital grade AC power source.
Caution—
grounding of the AC power source, operate the monitor only from its battery.
2
Verify that the Battery Charging Indicator LED is on. Allow the unit to fully charge before
operating. See Recharging the Battery on page 2-3 for more information.
3
Attach the appropriate blood pressure hose to the NIBP hose connector on the back of the
monitor.
Select the appropriate sized BP cuff and attach it to the hose. Sizes range from neonatal to
large adult and thigh. For more information about choosing a BP cuff, see
Chapter 5, "Monitoring Blood Pressure."
4
Connect the SpO
will have a positive lock on the cable when inserted correctly.
Select the appropriate SpO
5
Connect the appropriate temperature probe (oral or rectal) by connecting the temperature
probe cable to the temperature probe connector on the back of the monitor. Push the cable
connector in until a positive latch is made.
Note: For oral temperature measurements, use the blue probe; for rectal, use the red probe.
Insert the probe into the probe well and place a new box of probe covers into the probe
cover box receptacle.
2-2
Setting Up the Monitor
1
2
AC 100 ~ 240
50/60 Hz, 2V A
In tern ally Fu sed 2A
Use only the supplied power cord. If in doubt about the integrity of the
extension cable to the SpO
2
sensor and attach it to the extension cable.
2
Handle
Barcode Scanner I/O
Connector
Temperature Probe Connector
T
Rs232
SpO2
SpO
connector on the back of the monitor.You
2
3
1
1
Cable Connector
2
Label
100 - 240V
50/60 Hz
120VA max
SpO
2
T