5 Preparation for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
5.1 Installing the foot adapter
► Install the foot adapter on the prosthetic foot according to its instructions
for use.
5.2 Alignment
NOTICE
Grinding the prosthetic foot
Premature wear resulting from damage to the prosthetic foot
► Do not grind the prosthetic foot.
5.2.1 Bench Alignment
TT bench alignment
Recommended tools:
662M4 goniometer, 743S12 heel height measuring apparatus,
743A80 50:50 gauge, alignment apparatus (e.g. 743L200
L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly)
Perform the alignment and assembly of the prosthetic components in the
alignment apparatus according to the following specifications:
Sagittal plane
Heel height:
Effective heel height (shoe heel height - sole thickness of fore
foot) + 5 mm
External foot rotation:
approx. 5°
a–p positioning, middle of the prosthetic foot to the alignment refer
ence line:
Positioning according to the table below
Connect the prosthetic foot and prosthetic socket using the chosen
adapters.
Bench alignment process
13