Clinical measuring
Complies with the requirements of EN 1060-4;
accuracy:
Korotkoff validation method: Phase I (SYS),
Phase V (DIA)
Operating mode:
Continuous mode
Nominal voltage:
DC 3V
Expected service life:
10,000 measurements
Power supply:
2 x 1.5 V alkaline manganese (AAA/LR03)
batteries
Battery capacity:
Approx. 1,000 measurements
Protection against
Internally powered ME device Application
electric shock:
part: Type BF
Protection against
IP22 (not waterproof)
harmful ingress
of water or solid
materials:
Inflation pressure:
About 190 mmHg for first measurement
Automatic switch-
1 minute after measurement / otherwise 30
off:
sec.
Cuff:
For wrist diameters of
12.5 - 21 cm
Memory capacity:
2 x 100 measurements with average value
of all measurements and morning / evening
average value of the last 7 days
Operating
Ambient temperature: +10°C to +40°C
conditions:
Relative humidity:
≤ 85%, non-condensing
Air pressure: 800 - 1050 hPa
Storage / transport
Ambient temperature:
conditions:
–20°C to +50°C
Relative humidity: ≤ 85%, non-condensing
Serial number:
Inside the battery compartment
69
English
EN