Expected Service Life; Product Disposal - DeVilbiss DV6HH Serie Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 7
MAINTENANCE
Humidifier Water Chamber (single-patient use) - Clean the chamber
daily. The chamber should be replaced according to the schedule
recommended by your provider. Inspect the chamber monthly; replace if
damaged.
Humidifier Cradle - No maintenance needed for humidifier cradle.

EXPECTED SERVICE LIFE

Humidifier Cradle - 5 years
Humidifier PulseDose Valve (only model DV6HHPD) - 5 years
Humidifier Water Chamber - 1 year
TROUBLESHOOTING
Issue
Water condensation collecting in the
hose causing a gurgling sound
when using optional humidifier.
Heater setting is not visible on
display.
SPECIFICATIONS
Humidifier
Size with CPAP Device: ..................................................................................................................................................... 6.3" (16.1 cm) H x 6.1" (15.5 cm) W x 7.6" (19.3 cm) D
Weight: .................................................................................................................................................................................................................. 3.8 lbs (1.75 kg) w/CPAP device
Humidity Output (in the operating flow range): ................................................................................................................................................................................≥10 mgH
Maximum Power Consumption: ................................................................................................................................................................................................ 40 watts (humidifier)
Typical Power Consumption with CPAP ....................................................................................................................................................................................................... 25 watts
DV6HH Electrical Requirements: ........................................................................................................................................................................... 11-17 VDC powered from CPAP
Heater Plate Thermal Cutout:.............................................................................................................................................................................................................. 239˚F (115˚C)
Power/Temperature Control: ........................................................................................................................................................ Humidity Setting of 1 (minimum) to 5 (maximum)
Heater Plate Temperature: ...........................................................................................................................................................................................91˚F to 149˚F (33˚C to 65˚C)
Operating Temperature Range: ......................................................................................................................................................................................41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range: ................................................................................................................................................................................. 15% to 93% R.H. non-condensing
Operating Atmospheric Pressure Range: ....................................................................................................................................................................................... 70.0 – 106.0 kPA
Operating Flow Range: .................................................................................................................................................................................................................. 15 L/M to 60 L/M
Not manufactured with natural rubber latex.
EN - 6
Possible Cause
1. Humidification is too high.
2. Room temperature fluctuates from higher to lower levels
through the night.
1. The flow generator is not fully seated onto the humidifier cradle.

PRODUCT DISPOSAL

The outer packaging is made of environmentally friendly materials that can be used as secondary
raw materials. If you no longer need this packaging, bring it to your local recycling and waste
disposal facility according to the applicable regulations.
The equipment, including accessories and internal components, does not belong in your regular
household waste; such equipment is manufactured from high-grade materials and can be recycled
and reused. The PC board, heater plate, and wiring should be removed from the unit and recycled
as electronic waste. The remaining plastic components should be recycled as plastics.
The European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE)
requires that electrical and electronic equipment be collected and disposed of separately from
other unsorted municipal waste, with the aim of recycling it. The crossed out waste bin symbol
indicates that separate collection is required.
Remedy
1. Reduce humidifier heater setting.
2. Increase room temperature.
1. Refer to System Assembly to ensure proper contact is
made.
O/l air
2
A-DV6HH

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Dv6 serieDv6hh

Tabla de contenido