Reprocessing

Warning, electric shock!
Disconnect the device from the mains before replacing the fuses.

Disassembly

9.1
Disconnect the mains plug and the device plug
Disconnect the plug of the hose heater
Disconnect the hoses and the bacterial filter.
Push the holding-down device up and remove the lid of the nebulizer chamber. Detach the nebulizer
chamber and empty out the fluid. Observe the hospital hygiene regulations
Dispose of disposable articles (disposable hoses, medication container and lid, containers of sterile water
systems) in an environmentally-friendly manner. Observe the hygiene regulations.
Do not disinfect or sterilize the filter
Replace and dispose of the filter as directed by the manufacturer.
9.2

Notes on implementation

The individual parts of the ultrasonic nebulizer can be cleaned and disinfected according to the information in the
table below. Please observe the following information to do this:
-
The parts must be cleaned and dried before every disinfection.
-
Dirt particles can be encapsulated and the device parts will therefore not be sterile after disinfection. This is
why all device parts must be thoroughly cleaned and persistent soiling removed with a cloth and cleaning
agent.
-
Please observe the instructions of the manufacturer and the hygiene expert before using disinfectants.
Warning, risk of injury!
Disinfectants can contain substances that are harmful to health and cause injuries when they come into
contact with skin or eyes. Protect your skin and eyes and comply with the hygiene rules when working
with disinfectants.
Caution, material changes!
Almost all components of the ultrasonic nebulizer are made of plastics. Solvents, some disinfectants and
cleaning agents can corrode plastics or cause stress cracks. Do not use agents containing hydrocarbons or
alcohol to clean the surfaces.
Caution, functional inspection!
The functionality of the device parts must be checked after every cleaning or disinfection (7.1 Functional
test).
Note
If wrapping cloths that are not color fast are used, this can discolor the plastic parts.
Note
After every cleaning, disinfection and complete drying, check the rubber seal of the quartz module
thoroughly for potential damages.
9.3

Implementation

Components
1
Nebulizer chamber
(brown color)
1
Nebulizer chamber lid
(white)
1
Nebulizer chamber
with
quartz module
Hoses
1
Nebulizer chamber
(transparent)
96200-xxx_00 GA USV Tropic 2.0_20160229.docx
Cleaning
Disinfection
in warm water with
mild household
immerse in diluted solution
cleaner
in warm water with
immerse in diluted solution
mild household
Dishwasher
2
rinse cycle up to
95 °C
not allowed
rinse cycle up to
95 °C
Hot steam
3
sterilisation
up to 134 °C
18min
up to 121 °C
20 min
- 58 -