General electronic devices
Electromagnetic immunity
14.3
Guidelines and manufacturer's declarations – electromagnetic immunity
The ultrasonic nebulizer is intended for use in the electromagnetic environment specified below.
The customer or user of the product should ensure that it is operated in such an environment.
IEC 60601
Immunity test
Test level
Electrostatic discharge/
± 6 KV
ESD
contact discharge
according to IEC 61000-4-2
± 8 KV
air discharge
Electrical fast
± 2 KV
transients/bursts
for power supply lines
according to IEC 61000-4-5
Surges on AC mains lines
± 1 KV
according to IEC 61000-4-11
differential mode
voltages
± 2 KV
common mode voltages
Voltage dips, short
interruptions and
>95 % dip, 0.5 periods
fluctuations of the supply
60 % dip, 5 periods
voltage
30 % dip, 25 periods
according to IEC 61000-4-11
>95 % dip, 5 seconds
Power frequency magnetic
field
(50/60 Hz)
3 A/m
according to IEC 61000-4-8
Note: U
is the alternating mains voltage prior to application of test levels.
T
Electromagnetic immunity, devices not intended for life support
14.4
Guidelines and manufacturer's declarations – electromagnetic immunity, devices not intended for life support
The ultrasonic nebulizer is intended for use in the electromagnetic environment specified below.
The customer or user of the product should ensure that it is operated in such an environment.
Immunity test
Conducted RF disturbances
according to IEC 61000-4-6
Radiated RF disturbances
according to IEC 61000-4-3
96200-xxx_00 GA USV Tropic 2.0_20160229.docx
Malfunction
e.g. stop or change of the intended operation
mode
Compliance
level
± 6 KV
contact discharge
± 8 KV
air discharge
± 2 KV
for power supply lines
± 1 KV
differential mode
voltages
± 2 KV
common mode voltages
-100 %
-60 %
-30 %
-100 %
3 A/m
IEC 60601
Compliance
Test level
level
3 V
3 V
eff
eff
150 kHz – 80 MHz
150 kHz – 80 MHz
3 V/m
3 V/m
80 MHz – 2.5 GHz
80 MHz – 2.5 GHz
Electromagnetic environment –
guideline
Floors should be made from wood or concrete or tiled
with ceramic tiles.
If the floor is covered with synthetic material, the relative
humidity level must be at least 30 %.
Mains voltage quality should be that of a typical
commercial or hospital environment.
Mains voltage quality should be that of a typical
commercial or hospital environment.
Mains voltage quality should be that of a typical
commercial or hospital environment.
If the user of the ultrasonic nebulizer requires continued
operation even during energy supply interruptions, it is
recommended that the ultrasonic nebulizer is powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical commercial or hospital
environment.
Electromagnetic environment –
guideline
Portable and mobile radio equipment should not
be used at a smaller distance to the product,
including lines, than the recommended separation
distance, which is calculated by means of the
equation appropriate for the transmission
frequency.
Recommended separation distance:
d = 1.2 √P (for 150 kHz – 80 MHz)
d = 1.2 √P (for 80MHz – 800 MHz)
d = 2.4 √P (for 800 MHz – 2.5 GHz)
with P as nominal output of the transmitter in
watts (W) in accordance with the transmitter
- 63 -