Vercise™ DBS Remote Control 3 Directions for Use
Warning: An MRI scan is not recommended when the impedances are not within the acceptable
range. Higher or lower than normal impedances could indicate compromised Stimulator-Lead integrity.
Scanning under these conditions may increase the risk of potential adverse effects listed in the Safety
Information section of the ImageReady
Stimulator Error Screen
If the system check fails due to a Stimulator error, MRI Mode will not be enabled and the Remote
Control will display the Stimulator Error Screen. Do not perform an MRI scan if this error is displayed.
Instruct the patient to contact their physician managing their DBS System or Boston Scientific.
End of Programmed Service (For Non-rechargeable Stimulators Only)
The non-rechargeable Stimulator software has been programmed to end service after a defined period.
When the non-rechargeable Stimulator is within approximately 180 days of the end of its programmed
period, the Remote Control will display the following on-screen message.