ERI (Elective Replacement Indicator) Screen (For Non-rechargeable Stimulator Only)
Press
to clear the message. The message will reappear approximately weekly until the end of the
non-rechargeable Stimulator's programmed period is reached. When the end of its programmed period
is reached, the Remote Control will display the following on-screen message.
ERI (Elective Replacement Indicator) Screen (For Non-rechargeable
Stimulator Only)
This screen displays when the implanted non-rechargeable Stimulator is nearing end of service.
Changes made to the stimulation will not be saved, and stimulation will not be available soon. Batteries
that have lasted one year or more without entering ERI mode will have a minimum of four weeks
between entering ERI mode and reaching end of battery life. Patients should be advised to contact their
physician to report this message screen. Press
to clear this information screen. This screen will
appear whenever the Remote Control is activated.
Vercise™ DBS Remote Control 3 Directions for Use
91172931-02 REV A 29 of 835