7.2 - "Date" Menu
You can choose among 3 formats: DD/MM/YY, MM/DD/YY, YY/MM/DD.
¨
©
Press
or
Enter the date and time and press
7.3 - "Language" Menu
ª
«
Use the
and
A message is displayed to ask if this request should be made each time the unit is started.
7.4 - "Access Code" Menu
The access code allows you to protect the unit against any keying error.
On delivery, no code is programmed.
New
Conf.
|
The symbol
protection, enter the code; the symbol
delete the code, enter it again and press
7.5 - "Auto stop" Menu
You can opt or not for the validation of the automatic stop of the unit.
The time can last from 1 to 60 mn. It lasts 3 mn by default.
7.6 - "Buzzer" Menu
This menu allows you to validate or not the use of the buzzer in the measurement screen (to indicate a reject report).
8 - TACHOMETER
This screen allows you to measure the rotation speed of the transducer.
Since the transducer is that corresponding to the task in progress, you must select a torque and angle transducer* (of the
CMDxxxx** or DRT5xxx type). The message "Wrong transducer" is displayed if a torque transducer is selected.
The screen displays the speed in rpm and also displays the torque in "track" mode.
* DELTA5000/5D only
** DELTA4000/5000 only
9 - MAINTENANCE
9.1 - "Input/Output" Menu
This menu allows the operator to:
- know the status of the 2 inputs: Reset / Save and Reset / No save.
- check the control of report LEDs and outputs.
9.2 - "Transducer" Menu
It is used to reverse the reading direction of the transducer. This selection is due to various types of CNA cable
versions.
DELTA4000/5000/4D/5D
to select the date format. Press
¢
to validate or
keys to select the language and press
Enter the new code (8 alphanumerical characters as a maximum).
Confirm by entering the code a second time, then press
is displayed and all the calculation parameters become available in read-only mode. To unlock the
}
is displayed. You can then update the parameters and lock them again. To
¢
to validate; the symbol
¢
to validate.
¡
to exit the screen without saving changes.
¢
to validate.
¢
|
is no longer displayed.
6159938010-07
English
15 / 200
to validate.