Browsing aid in the operating mode:
Type of joint
CVIC Calibration
Note: the "CVIC calibration" section is described in the CVIC Operator's Manual.
In calibration mode, select the type of transducer which corresponds to the CVIC application.
Automatic printing of the results:
¥
Press
to display the request screen for the automatic printing of the measurements:
Enable
yes/no (auto print or not)
Periph.
Selection of the peripheral: (parallel) printer or PC (serial link).
Header
yes/no (printing without header or footer).
Format
Standard/Specific
The "automatic printing" function allows the operator to print the results as and when the measurements are collected. To exit
the "Acquisition" menu, press
Transducer connection:
The transducers that may be connected to this measuring unit are the CMD** series analog transducers or the DRT series
digital transducers. DRT type transducers are automatically recognised by the measuring unit, and their characteristics are
automatically taken into account.
The selection of a transducer type is determined by the selection of a task on the measurement screen; several cases can be
considered if the transducer determined by the task is not identical to the transducer actually connected.
Transducer connected
DRT4xx or DRT5xx
Analog transducer (no digital
transducer detected)
Analog transducer (no digital
transducer detected)
Digital transducer DRTxx
Conformity between the task-oriented transducer and the transducer actually connected, is continuously checked on the
measurement screen, in order to take account of a change even if made with the transducer switched on.
** DELTA4000/5000 only
4 - "PARAMETERS" MENU
To exit the "Acquisition" menu, press
This menu allows the operator to program the parameters of the task, parallel printer and serial link.
You can also request the printing of all the parameters that you have selected: press
the "Parameters" menu. In the latter case, you can request the printing of every task, the parameters of the serial link, of the
printer, and of the curves.
Note: The unit can be reset (erasing all parameters and measurements) by pressing
DELTA4000/5000/4D/5D
or
Track
Peak
etc...
¡
.
Task-oriented transducer
Other DRT transducer or
analog transducer of CMDxx
type, etc.
DRTxx
Analog transducer
Same transducer DRTxx
¡
. The main menu is displayed.
Action
The transducer normally provided for this task is replaced by
the transducer connected, after validation by the operator.
Prohibit measurement
Locking warning message (no transducer connected)
OK
OK
¥
on the task modification screen or in
¥
when switching on.