Transportation and installation
Transport and unpacking
4.1
All products from Dentsply Sirona are carefully checked prior to
shipment. Please perform an incoming inspection immediately after
delivery.
1. Check the delivery note to ensure that the consignment is complete.
2. Check whether the product shows any visible signs of damage.
NOTE
Damage during transport
If the product was damaged during transport, please contact your
carrying agent.
If return shipment is required, please use the original packaging for
shipment.
Disposal of packaging materials
4.2
The packaging must be disposed of in compliance with the relevant
national regulations. Please observe the regulations applicable in your
country.
Installation conditions
4.3
4.3.1
Installation site requirements
The inLab MC X5 suction device requires an even base of approx.
400 mm x 400 mm (W x D).
The CEREC MC/MC X/MC XL Premium Package/Primemill suction
device requires an even base of approx. 360 mm x 260 mm (W x D).
Install the suction device in such a way that the main switch can be
pressed easily. To enable simple replacement of the filter and bag, we
recommend orientating the maintenance flap to the front with the two
knurled nuts. The ventilation slots on the bottom side and rear side of
the suction system must be kept unobstructed. The distance between
the back of the unit and the wall must at least be 30 mm. If the suction
device is integrated in a cabinet, this must be vented sufficiently to
prevent the suction device from overheating.
The suction device must not be set up in the immediate vicinity of
considerable humidity development.