7 Malfunctions

7.2 Troubleshooting and repair

Symptom
Suction device does not start up, ei-
ther in automatic mode or with the
on/off button set to On.
Insufficient suction power
Very strong suction power
Suction very loud
24
Troubleshooting and repair
7.2
Possible cause
Power cable not connected.
Voltage supply inactive.
The fuse on the suction device has
tripped.
Suction motor superheated.
Suction motor defective / worn out.
Suction electronics defective.
Suction power controller is at min
on the control panel.
Filter bag is full.
HEPA filter occupied.
Suction tube blocked.
Leak on suction tube or socket.
Filter bag / HEPA filter not seated
correctly.
Turbine touches the housing.
Dentsply Sirona
Operating Instructions for suction devices
Remedy:
Check power cable for correct con-
nection.
Check voltage source
Press the device safety switch, see
Technical description (component
and interfaces)  [→ 12]
Let the suction system cool down;
ensure there is sufficient air circula-
tion, especially around the ventila-
tion slots.
Contact services
Contact services
Set suction power controller to max.
Replace filter bag
Replace HEPA filter
Check suction tube for open pas-
sage; remove blockages if neces-
sary.
Check suction tube for leak tight-
ness and tight connection to the
suction unit and processing ma-
chine; seal the tube / connect it
tightly if required.
Open maintenance cover. Check fil-
ter bag and HEPA filter are seated
correctly.
Contact services
64 98 609 D3608