Accuracy Table
(3) AC INRUSH (Rush current)
AC INRUSH measurement value (rms)
Accuracy guarantee range
Range
60.00 A
3.00 A to 60.00 A
(0.01 A)
600.0 A
10.0 A to 600.0 A
(0.1 A)
2000 A
100 A to 2000 A
(1 A)
*1: Design values apply beyond 6×10
AC INRUSH PEAK value (Zero to Peak)
Accuracy guarantee range
Range
60.00 A
3.0 A to 150.0 A
(0.1 A)
600.0 A
10 A to 1500 A
(1 A)
2000 A
100 A to 2840 A
(10 A)
46
(Resolution)
5
A · Hz.
(Resolution)
Accuracy guarantee
frequency range
40 Hz ≤ f ≤ 500 Hz
40 Hz ≤ f ≤ 500 Hz
1
40 Hz ≤ f ≤ 500 Hz*
Accuracy guarantee
frequency range
40 Hz ≤ f ≤ 500 Hz
40 Hz ≤ f ≤ 500 Hz
40 Hz ≤ f ≤ 500 Hz
Measurement accuracy
±5.0% rdg.±0.13 A
±5.0% rdg.±1.3 A
±5.0% rdg.±13 A
Measurement accuracy
±5.0% rdg.±1.0 A
±5.0% rdg.±10 A
±5.0% rdg.±100 A