(12) Capacitance
Range (Auto-range threshold)
1.000 µF
(more than 1100 counts)
10.00 µF
(more than 1100 counts/less than
100 counts)
100.0 µF
(more than 1100 counts/less than
100 counts)
1000 µF
(less than 100 counts)
(13) Temperature (Thermocouples (K))
Range
°C
40.0°C to 400.0°C (0.1°C)
2
°F *
40.0°F to 752.0°F (0.1°F)
*1: Conditions (In an environment where the temperature of the instrument is ±1°C and stable)
*2: Instrument can be made to display readings in Fahrenheit (°F) by means of special operation.
Accuracy guarantee range
(Resolution)
0.000 µF to 1.100 µF
(0.001 µF)
0.00 µF to 11.00 µF
(0.01 µF)
0.0 µF to 110.0 µF
(0.1 µF)
0 µF to 1100 µF
(1 µF)
Accuracy guarantee range (Resolution)
Accuracy Table
Measurement
Discharge current
accuracy
10 nA±20%
±1.9% rdg.
100 nA±20%
±0.005 µF
1 µA±20%
100 nA±20%
±1.9% rdg.
1 µA±20%
±0.05 µF
10 µA±20%
1 µA±20%
±1.9% rdg.
10 µA±20%
±0.5 µF
100 µA±20%
10 µA±20%
±1.9% rdg.
100 µA±20%
±5 µF
200 µA±20%
Measurement accuracy *
±0.5% rdg.±3.0°C
±0.5% rdg.±5.4°F
3
1
57