Current Measurement
Auto hold conditions
Display value updates are stopped when the following two conditions are satisfied:
• When the measured value exceeds the threshold value described in the table in the next page. (voltage,
current). When the measured value is less than the threshold value described in the table in the next page.
(resistance, continuity, diode)
• When the range over which the measured value is fluctuating stabilizes within the fluctuation range
described in the table in the next page.
Held measured value
If the measured value falls below the threshold value (voltage, current) or exceeds the threshold value
(resistance, continuity, diode) after display value updates are stopped, the display value update is restarted.
Display value updates will stop if the two conditions are satisfied once again.
16
Auto hold
e.g.: 100.0 A
Fluctuation range
Measured
value
Fluctuation range
End
Start
Start
Auto hold
e.g.: 99.0 A
Fluctuation range
End
End
Start
Measurement
Threshold
e.g.: 12.0 A
Time