USER'S MANUAL. PROLITE-40
10 APPENDIX A: PRACTICAL USE OF THE SPLICER
PROLITE-40
This addendum can help you to resolve some questions by yourself.
The screen is dark after starting machine.
Solutions:
1.- Check whether the electrical outlet is inserted.
2.- Check whether the voltage of P.S. is too low.
3.- Check electric quantity of battery and charge battery in time.
"Exhausted Battery" is displayed on the bottom of screen after starting
PROLITE-40 and buzzer works continually.
Solutions:
1.- If use battery, charge in time or replace the power supply.
The fiber can be reset as normal, but when setting gap, fiber is immovable on the
screen and "Setting gap" is displayed on the screen.
Solutions:
1.- Raise the windshield, left and right clamps. The checking steps as follows.
2.- Check whether broken fiber exists.
3.- Check whether cleaved length is too short.
4.- Check whether fiber matches with V-groove.
5.- Check whether fiber is tail fiber.
The fiber can be reset as normal, but when setting gap, fiber moves backwards
continually, "Setting gap" and "Reload fiber" are displayed on screen.
Solutions:
Maybe there is too much dust on the lens of microscope, clean the a clinic and
upright lens with cotton swab until dust disappears, then try again.