DO NOT touch the sampling end of the test strip with the bottle of control
solution. See illustration of sampling end below.
Step 5:
Gently touch the tip of the test strip to the drop of control solution.
The meter will beep indicating it has received the control solution.
The screen will start to count down. After 6 seconds, the control
solution test result will appear on the meter display screen.
Step 6:
Compare the reading on the screen to the Low range printed
on the test strip bottle. If the reading falls outside the Low range
printed on the test strip bottle, see Control Solution
Troubleshooting, on page 31.
NOTE:
• Control solution test results will be stored in meter
memory and indicated by ctl icon.
• Control solution test results will not be used for
calculating averages.
Step 7:
Remove the used test strip. The meter will automatically turn off.
Discard the used test strip. Follow steps 8 through 14 to perform the high
control solution testing.
IMPORTANT:
Do not reuse test strips.
Control Solution Testing
G
EVENCARE
®
2
2
EVENCARE G
Low
53∼79
High
135∼203
Low
High
29