Step 8:
Take out an EVENCARE G2 Test Strip from the
test strip bottle and close the bottle immediately.
Insert the test strip to turn on the meter.
Step 9:
Wait until the flashing blood drop and arrow icons
appear on the meter display screen. The meter will
also announce, "Please apply blood onto the test
strip" if you have the voice feature turned on.
Press
two times to enter L2 (HIGH) control
t
solution testing.
Step 10:
ctl icon will appear next to the test strip icon and
L2 will appear on the meter display screen. The
meter will also announce "L2 control solution test,
please apply control solution onto the test strip."
If you decide not to do a control solution test, press
one time and ctl will disappear from the meter display.
t
Step 11:
Squeeze a drop of High control solution onto a clean,
dry, non-absorbent surface. Do not apply control solution
to the test strip directly from the control solution bottle.
Replace the bottle cap on the control solution bottle
immediately after use.
Step 12:
Gently touch the tip of the test strip to the drop of
control solution. The meter will beep indicating it has
received the control solution. The screen will start to
count down. After 6 seconds, the control solution
test result will appear on the meter display screen.
30
Control Solution Testing
mg/dL