Step 13:
Compare the reading on the screen to the High
range printed on the test strip bottle. If the reading
falls outside the High range printed on the test strip
bottle, see Control Solution Troubleshooting below.
Step 14:
Remove the used test strip. The meter will
automatically turn off. Discard the used strip.

Control Solution Troubleshooting

If your control solution result is out of range (too high or too low), it may
be caused by the following:
Possible Causes
• Wrong brand of
control solution
being used.
• Control solution
not at room
temperature.
• Expired or
contaminated
control solution,
or damaged test
strip.
• Meter malfunction.
Control Solution Testing
What you can do...
• Make sure you are using EVENCARE G2
Control Solution.
• Make sure the testing environment is
between 50˚F to 104˚F.
• Check the expiration and open bottle date
on both the control solution and the test
strips. Repeat the test using a new test
strip. If the result is still out of range, use
a new bottle of control solution and retest.
• If the problem persists, please call
Medline's Technical Service Center at 1-866-
274-2131 Monday through Friday between
8:00 a.m. and 5:00 p.m.
(Central Time).
2
EVENCARE G
Low
53∼79
High
135∼203
Low
High
31