Calibration Errors - Adam Equipment Nimbus NBL Serie Manual De Instrucciones

Ocultar thumbs Ver también para Nimbus NBL Serie:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

 Internal temperature change greater than a pre-set amount (typically 2ºC for
Precision balances).
 Time since last calibration exceeds a pre-set time (typically 4 hours, or 15 minutes
after power is applied).
The balance will indicate the need for calibration to be carried out by flashing the "
symbol on the display. As soon as the balance is calibrated the symbol will be turned off.
The Auto calibration feature can be enabled, disabled or changed within the user options
to meet the requirements of the users.

6.2.5 Calibration Errors

Occasionally during calibration an error will be detected. These errors can be caused by:
 Unstable readings
 Improper calibration weights being used
 Large shifts of zero from the factory settings
When an error is found a displayed message will be shown and the calibration must be
done again. If the balance has error messages more than once it is possible the
mechanics have been damaged.
© Adam Equipment Company 2017
CAL
"
15