Select a code for one of the pre-set data formats as detailed above.
The next line shows: "" "" - prints time.
Only one item can be entered per line.
Continue until the formatting of the form is complete. There are 15 lines of possible data.
After the 15th line has been set or "" has been selected, the balance will return to the
"" Sub-menu.
Press [Mode] to advance to setting of the next menu, or press [0/T] to return to
normal weighing.
13.5 SETUP PARAMETERS
The user parameters that control the balance are shown under the setup menu. When
"" is displayed, press the [Setup] key. The options for each parameter can be
scrolled through by using the [Up] or [Down] key. Use the [Up] and [Down] keys to
increase or decrease the value for setting. Press [Setup] to accept the setting and
advance to the next item in the menu
Press [Mode] to advance to setting of the next parameter or [0/T] to return to normal
weighing
Select menu language from available options.
Set real-time clock using the keypad numeric entry method. HH:MM:SS.
Set date display format using the keypad numeric entry method. European
(DD/MM/YY) or USA format (MM/DD/YY).
Set date using the keypad numeric entry method. YEAR, MONTH, DAY,
WEEKDAY
Enter a user number to identify this balance on print output. Range 1 -
9999999
On = Enable sound alerts
OFF = Disable sound alerts
AUTO = Always on unless balance is not used for 5 minutes, then turns off
automatically until key is pressed or weight >20d is detected.
ON
OFF = Permanently off
On = Power-saving mode enabled. Sets the inactivity period after which unit
will go into stand-by mode. Range 1 – 9 minutes.
OFF = Power-saving mode disabled.
The filter tracks and averages weighing to produce the most accurate
measurement and smooth out instabilities. A higher filter number means more
filtering and a slower, but possibly more stable and accurate response. A
lower number will produce a quicker measurement but it may be less stable
and accurate. Range 1 (low) to 9 (high).
Recommended value for normal use: 5
ON = A fine filter which provides better performance when weighing whilst
pouring a substance such as liquid or powder into a container on the pan.
OFF = No filtering. Recommended setting for normal use.
Set a value to be used to determine balance stability. The number
corresponds to the number of divisions the weight reading is fluctuating by. A
© Adam Equipment Company 2017
= Permanently on
36