Operating Guidance - KPS MT60 Manual De Funcionamiento Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

AC Current
Range
Resolution
660uA
0,1uA
6600uA
66mA
10uA
660mA
100uA
10A
10mA
- Overload protection:
- A and mA grade: Fuse FF600mA/1000V, A grade: Fuse FF10A/500V.
- Frequency range: 50 ~ 60Hz
- Response: true RMS
When measuring current is greater than 5A, the continuous measurement time
should not be more than 10 seconds. After measuring, disconnect the current.
4.

OPERATING GUIDANCE

4.1 Reading hold
• In the process of measurement, if reading hold is required, press "HOLD"
key, the value on the display will be locked.
• Press "HOLD" key again to cancel reading hold.
4.2 Measuring range switch
• When turning the conversion switch to current, voltage, resistance, capaci-
tance, frequency grade, the meter is in automatic measuring range mode.
• Press "RANGE" key, the meter will enter the manual range mode. Press
once, the measuring range will go up with one grade. If it is pressed at the
highest measuring range, it will go to the minimum measuring range.
• If the user presses "RANGE" key more than 1 sec, the meter will restore
automatic measuring range.
• If the meter is in the maximum/minimum measurement state, pressing the
"RANGE" key will return the meter to normal measurement state.
1uA
KPS-MT60 • Digital multimeter
Accuracy
±(1,5% of reading + 5 digits)
±(3,0% of reading+ 5 digits)
32