Removing E-type handpieces (According to ISO 3964 type)
1) Open the door after the Connector LED has turned OFF and lubrication has completely
finished.
2) Pull out the handpiece straight while pressing the Release Button on the E-type Handpiece
Connector (Fig. 19).
• With a cloth, wipe off any Oil collected at the bottom of the chamber and the
door. The collected oil could leak from the bottom of the door. Dispose of the
CAUTION
cloth as medical waste according to the laws and regulations in your country.
8-4 Chuck Cleaning
This product can clean the Chuck of Air Turbine and Contra Angle handpieces.
1) Insert the chuck cleaning nozzle into the Connector "4" until it locks (Fig. 20).
2) Press the Chuck Cleaning Button to illuminate the Chuck Cleaning LED.
3) Holding the handpiece securely by hand, insert the chuck cleaning nozzle into the bur insertion hole of the handpiece (Fig. 21).
4) Press the Start Button and chuck cleaning will start. The Chuck Cleaning LED will blink during cleaning.
5) When cleaning is completed, a beep sounds and the Chuck Cleaning LED will turn off.
6) Lubricate the handpiece following the instruction in "8-1 Attaching Handpiece" and "8-2 Lubrication and internal cleaning".
• Do not clean the chuck of NSK Straight Handpieces as this may cause handpiece malfunction.
CAUTION
NOTICE
• Only Connector "4" is available for chuck cleaning.
• Clean the chuck at least once a week.
9

Maintenance of iCare

9-1 Cleaning the iCare
1) Turn OFF (O side) the Power Switch.
2) Remove the AC power cord from Unit.
3) Wipe the surface using a cloth moistened with water, warm water or soapy water (neutral detergent), then again wipe using another cloth
moistened with disinfectant alcohol.
• Never use solvent such as benzine or thinner for cleaning.
• Do not leave an alcohol-immersed cloth in contact with the Unit for a long period. Bleaching, deformation or failure of the Unit
CAUTION
may occur.
NOTICE
• The door can easily be detached when necessary and be cleaned with tap water (See "9-2 Detaching and reattaching the
door").
E0643E002 (CS5) iCare-EN_141001.indd
11
Chuck Cleaning Nozzle
*C2 Type
11
Release Button
*C2 Type
Fig. 20
Fig. 19
Fig. 21
2014/10/01