ProMinent DULCOMETER DMT Instructions De Service página 21

Ocultar thumbs Ver también para DULCOMETER DMT:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

SensorType
SensorQuality
Item Name
0
2
2
1
2
2
2
3
4
2
5
2
6
2
-2
DMT-Redox (ORP)
MessStatus
Item Name
2
0
1
2
2
2
3
2
4
2
2
5
6
2
2
7
SensorType – no significance
SensorQuality
Item Name
0
2
-2
4
2
5
2
6
2
-2
ProMinent
®
Name
With sensor monitor
Without sensor monitor
Min. slope limit
Max. slope limit
Min. zero point limit
Max. zero point limit
Setting time
Probe outside
calibration rang
7
Measured value
Temperature correction
Value above max. limit
Value below min. limit
Sensor defective
3
Probe resistance
Probe outside
calibration range
7
1
Status 0
> 45 mV/pH (OK)
< 63 mV/pH (OK)
> -50 mV (OK)
< 50 mV (OK)
OK (< 60 s)
Last
calibration OK
Always
Status 0
Valid
ON
Value not above
max. limit
Value not below
min. limit
Sensor OK
Always
Always
Always
Status 0
Always
OK (> 2 MΩ)
Probe OK
Always
Functions
Value
0
Status 1
40 ÷ 45 mV/pH
63 ÷ 65 mV/pH
-60 ÷ -50 mV
50 ÷ 60 mV
Too long (> 60 s)
Calibration not
possible
Status 1
Invalid
OFF
MESS_VALUE>
MESS_MAX
MESS_VALUE<
MESS_MIN
Sensor resistance
<2 MΩ
Status 1
Probe defective
(<2 MΩ)
>40mV difference to buffer
solution
21

Publicidad

Tabla de contenido
loading

Tabla de contenido