NOTE: See FIGURE 3-1 for details
•
If If the "√" does not appear and and "ERR" is indicated:
• Check that the:
•
sensor inlet is not clogged
•
correct calibration cylinder is used to perform the bump
test
•
gas cylinder has not expired and is not empty
•
gas was applied at the appropriate time
•
gas tubing is seated in the front instrument case during
testing.
• Repeat the bump test process as needed.
•
If the "√" does not appear,calibrate the instrument and repeat the
Bump Test.
Figure 3-1. Bump Test Flow Chart