Warning; Calibrating The Altair Pro Single Gas Detector - MSA ALTAIR Pro Manual De Operación

Ocultar thumbs Ver también para ALTAIR Pro:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Chapter 4,
Calibrating the ALTAIR Pro Single Gas
Detector
Regulators and tubing used for bump testing and calibrating
Cl 2 , ClO 2 , NH 3 , HCN or PH 3 should be labeled by the user
for that specific gas and must only be used for that gas
exclusively in the future.
The ALTAIR Pro Single Gas detector must be calibrated if it does
not pass a Bump Test or if calibration is required by local
procedures. In addition:
For oxygen instruments, perform a calibration if:
there are changes in barometric pressure
(altitude changes)
there are extreme changes in ambient temperature and
humidity (see TABLE 6-2, "Instrument Specifications").
the instrument does not pass a bump test.
Calibration is required at certain intervals by local
procedures.
Special requirements for ALTAIR PRO O 2 -R Remote
Oxygen Detector:
Operate this instrument only in ambient
temperatures ranging from 0ºC to 40ºC
(32ºF to 104ºF).
If the instrument is taken into an environment where
the ambient temperature is 15ºC (59ºF) or more,
higher or lower, than its previous location, do not
use the instrument for a period of 40 minutes to
allow the sensor to adjust to the new conditions.
After this 40-minute period, perform a Fresh Air Set-
up (FAS) to re-zero the sensor. Perform the FAS in
fresh air only.
Quick changes in barometric pressure result in an
offset in the sensor reading for a few minutes. If this
occurs, do not use the instrument for a period of five
minutes; then, perform a Fresh Air Set-up (FAS) to
re-zero the sensor. Perform the FAS in fresh air only.

" WARNING

4-1

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Altair pro

Tabla de contenido