key to turn on the scale, In case display will show any error message or incorrect measurement, repeat the calibration again.

9. ERROR CODES

ERROR MESSAGE

-----
ERR 1
ERR 2
ERR 4
ERR 5
ERR 6
ERR 7
ERR 8
ERR 9
ERR 10
ERR 11
ERR 12
ERR 13
ERR 14
ERR 15
ERR 16
ERR 17
ERR 18
ERR19
ERR P
ERR L
--OL--
--LO--
FAI L H
/
FAI L L
/
FAI L
BA LO
/
LO BA
DESCRIPTION
Maximum load exceeded
Incorrect date
Incorrect time
Zero setting error
Key board error
A/D value out of range
Percentage Error
Calibration weight error
Unstable Reading
Wireless communiation failure
Communication protocol error
Accumulation error
Lack of unit weight
Lack of sample
Gravity error
Paper error
Tare out of range
Pre-tare error
Initialize zero error
Printer error
Approval setting error
Over range
Underload
Calibration Error
Battery low
USER'S GUIDE SL
SOLUTION
Unload or reduce weight
Enter the date by using format "yy;mm:dd"
Enter the time by using format "hh:mm:ss"
Zero setting range exceeded due to switching
on.(4%max). Make sure platform empty.
Check the keys and connecter.
Make sure platform empty and check the pan is installed
proper. Check the load cell connectors.
Please check input data, must be > 0.5d
Check the test weights for calibration or linear calibration
Check any air variation, vibration, RF noise and
touching some where.
Check the load cell and connecters.
Check wireless settings or change Com settings from the
wireless
Check communication settings
Max accumulation times 99 / 999 / weight 999999
Check unit weight entry data, must be >0.5d
Check counting samples entry, must be >20d
Check the gravity settings.
Gravity range must be 0.9xx ~ 1.0xx
Check the printer paper
Minus weight or overload.
Remove the load and restart scale again.
Check the pre-tare value
Calibration the scale.
Check the printer settings or connections
Check the PCB jumper settings. Must be connect
jumper pin to K1
Remove the load.
Re calibrate
Minus weight, check the platform and restart or
calibrate.
Check the test weights & Re calibrate
Re charge battery, check the voltages.