Error code
Err. 3
Err. 4
Err. 10
For all other error codes:
•
Note error code
•
Have the device repaired.
•
Quote the error code to the repair service
The following information is intended to
help purchasers ensure safe operation of
this appliance in their laboratory.
Instruct operators on the appli-
cation, possible hazards during
use and operation of this appli-
ance based on the user informa-
tion.
Ensure that this user information is avail-
able to operators.
EN
Cause
•
The vacuum is built-up
too quickly due to the
fact that the suction
system is blocked.
•
Motor fault.
•
Too much material
•
Mixing time too long,
the material has already
begun to set.
Information for the purchaser
•
Have filter cleaned or changed
(Sec. 3.3).
•
Ventilate bowl and remove (see
Sec. 2.6.1, points no. 2 and 3)
•
Have the device repaired.
•
Ventilate bowl and remove (see
section 2.6.1, points no. 2 and 3)
•
Only fill the bowl to the maximum
fill level.
This limit is valid for powders and
liquids measured before mixing.
•
Ventilate bowl and remove (see
section 2.6.1, points no. 2 and 3)
•
Select shorter mixing times.
A. Application
A.1 Correct use of the appliance
The Twister / Twister venturi vacuum
mixer is designed solely for homogene-
ous, bubble-free mixing of dental casting
and modelling materials such as plasters,
investments, and silicones.
- 10 -
Solution