Shutting valve
(illus. 9-10)
These valves are used to make independent from the
others each surgery connected to the same plant.
They are controlled by a 24 V signal coming from the
tip support and (only in case of dry systems) from the
probes of the Mini-Separator.
Draw. 9
AMALGAM SEPARATORS
Draw. 10
Amalgam separator-tank
(draw. 11)
The dimensions of the standard product allow its use
only in dry systems. It is advisable to install it in the
water unit or in the floor-box of the dental unit. If there
is no room enough it is convenient to install the Maxi-
Canister plus box fitted with the purposely designed
bracket. The amalgam separator can collect the liq-
uids coming from the spittoon provided that the wash-
ing is timed for 20 sec. max.
Draw. 11
Hydroseparator
(draw. 12a-12b-13-14-14b pag. 14-15)
Coupling the Hydrocyclone to the Mini-Separator we have obtained the Hydroseparator,
draw. 12a-13, the amalgam separator for dry systems, TÜV tested for a separation level
above 95%, with a flow of 3.5 l/min (DIN version) and with a flow of 5.5 l/min (ISO
version). The Hydroseparator can be installed in the water unit, or, fitted with its box,
at the bottom of the chair.
In case of semi-dry systems, one or more Hydrocyclones, draw. 12b-14, depending on
the size of the plant, should be installed in the engine room; also in semi-dry systems,
the spittoon draining unit draw. 14b (which is always fitted in dry systems
together with Hydroseparator) must be installed in the water unit of each dental
unit, so that liquids coming from spittoons will be conveyed directly into the
aspiration pipeline.
Draw. 12b
Draw. 14
Draw. 12a
Draw. 13a