RS232 Interface
CONNECTION
The RS232 Interface Kit cable terminates with a 3 pin Sub-D female connector.
Active pins: Pin 1 = Ground, Pin 2 = RXD, Pin 3 = TXD
The Interface connector will connect directly to a computer, an Ohaus printer and many other printers
Note: When the second RS232 interface has been installed, an icon will appear on the display:
OPERATION
RS232 OUTPUT
With the interface installed the terminal will operate in one of the following two ways:
Auto Print = off, Stable = on or off
Pressing PRINT will send the display data.
If Stable is set to On, the balance will wait for a stable reading before sending the stable data.
Auto Print = not off, Stable = on or off
The balance will automatically send data based on the settings in the Menu.
If Stable is set to On, only stable values will be sent.
EN-3