EN-4
Data sent from the Interface is in standard ASCII format terminated with a [CRLF] (carriage return-
line feed). The output format is as follows:
Printout string for g, kg, lb, oz units:
[weight]
[space]
[unit]
[space]
[stability indicator]
[space]
[G/N]
[space]
[
]
Miscellaneous
[
]
Term. Char(s)
Printout string print format for lb:oz unit
[weight]
[space]
[unit]
[space]
[weight]
[space]
[unit]
[space]
[stability indicator]
[space]
[G/N]
[space]
[
]
Miscellaneous
[
]
Term. Char(s)
Output example:
Weigh Mode
Printout
11.11 kg NET
12.34 kg G
11.11 kg NET
1.23 kg T
MODE: WEIGH
<no line printed>
9 characters (right justified)
1 character
3 characters max (left justified)
1 character
1 character
"?" when unstable, blank when stable
1 character
1 character
1 character
5 character
2 character
4 characters (right justified)
1 character
2 characters max (left justified)
1 character
7 characters (right justified)
1 character
2 characters max (left justified)
1 character
1 character
"?" when unstable, blank when stable
1 character
1 character
1 character
5 character
2 character
Description
Result line
If Printx Content Result is ON
If Printx Content Gross is ON and a tare value is
Gross value line
entered
If Printx Content Net is ON and a tare value is
Net value line
entered
If Printx Content Tare is ON and a tare value is
Tare value line
entered
Mode line
If Printx Content Application Mode is ON
Information line
If Printx Content Into is ON
Comment
RS232 Interface