NSK iPex II Manual De Operación página 4

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

If the battery fluid leaks and gets into your eyes, immediately flush your eyes
with water and consult with a medical specialist as it may cause vision loss or
blindness.
If the battery fluid leaks and adheres to your skin or clothing, immediately flush
it out with water as it may cause damage to your skin.
Conduct a maintenance check after each patient, following the instructions in '7.
Operation check using the Tester'. If all the bar graphs are not displayed during
the check, root canal length will not be correctly measured. In this case stop
using immediately and contact your Authorized NSK Dealer for repair.
The numerical values displayed on the LCD panel do not indicate the actual
distance from the end of the root canal. Only use them as guidance for
measurement.
Do not keep using the product with the battery indicator "
Normal operation or indication may not be performed.
Should the product function abnormally during operation, cease operation
immediately.
Do not use in combination with other equipments while the lip hook is attached
to the patient. The product may not perform correct measurement.
Be sure to prevent the lip hook, file clip and their connector parts from getting
contact with household power supply sources (such as electric outlets) as it
may cause electric shock.
Be sure to sterilize the file clip and lip hook by autoclave sterilization after each
patient.
Do not operate close to patients with cardiac pacemakers as there is a danger
that it may affect the pacemaker.
Do not use the product by connecting or integrating into other medical devices.
Keep away from explosive substances and flammable materials.
Do not allow any impact on to the product. Do not drop the product. Personal
injury or damage to the product may result.
This product is designed for use with dry cell batteries only. Use commercially
available dry cell batteries specified in this operation manual. Carefully read
the instruction manual of the dry cell batteries prior to use.
If the probe plug is not sufficiently inserted, measurement may not be
performed.
Avoid chemical solution for treatment on the lip hook or file clip. Use with the
solution adhered to the instrument may cause inflammation.
CAUTION
" flashing.
3

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido