SECTION VI: SPECIFICATIONS
Maximum input power (laser setting)
Recommended minimum working distance between fiber tip and tissue
Medical grade helium input pressure
Gas flow at recommended pressure
Fiber Length
Nominal spot size at fiber tip
Beam divergence
Maximum outer diameter of the distal portion of the fiber assembly
COMPATIBLE LASER SYSTEMS AND SURGICAL INSTRUMENTS
OmniGuide fibers are only intended to be used with OmniGuide handpieces and lasers. Please see the enclosed fiber compatibility chart for OmniGuide fiber/handpiece
compatibility. Please note that this chart may have changed between when you purchased from OmniGuide, and when you use your fiber or handpiece. If there is any discrepancy,
the most up-to-date version of the fiber compatibility chart can be found at omniguidesurgical.com/fibercompatibility.
DIRECTIONS FOR USE
See laser system operating manual for detailed instructions, maintenance, and
additional warnings and precautions.
For sterile cases, use of a Gas Filter Unit (GFU) is required. See GFU instructions for
details.
Inspect fiber packaging. Do not use if package is damaged or opened.
Y
Note: See product packaging for expiration date.
1. Use sterile technique to open the fiber package.
Handle the fiber with care. Do not bend it excessively (radius <2 cm) to avoid damage and
Y
increased possibility of failure during use
2. Insert the fiber into a compatible surgical instrument
Note: See surgical instrument instructions for use for detailed instructions, additional warnings
and precautions.
3. Attach the fiber connector to the laser adapter. Ensure fiber connector is securely locked to the
adapter. Verify gas is flowing from the fiber tip.
FOR FURTHER INFORMATION
If further information on this product is needed, please contact OmniGuide Customer Service at (888) OMNI-GUIDE (888-666-4484) in the US, or your authorized distributor.
See www.omni-guide.com/patents for information about the coverage of this OmniGuide® product by U.S. and/or foreign patent rights.
USA
M
OmniGuide, Inc.
4 Maguire Rd,
Lexington, MA 02421
Phone: +1 (617) 551-8444 | (888) OMNI-GUIDE
Fax: +1 (617) 551-8445
Email: sales@omni-guide.com | URL: www.omni-guide.com
SPECIFICATIONS
9 W (single pulse mode) / Maximum pulse duration of 100 msec / Pause for a minimum of 3
seconds between pulses
2 mm
See the Fiber Procedural Setting Guide (DOC-GEN-3) or the OR Staff Training Binder (DOC-
FELS-27) for recommended power and gas ranges.
< 1 lpm
SPECIFICATIONS
1.5 m
0.25 mm
200 mrad (11.46
, full angle)
0
0.55 mm
4. Turn on gas supply and set gas pressure.
5. Activate the laser energy to cut, ablate, or coagulate soft tissue for the desired application.
Do not use the fiber outside its recommended parameters in the Specification Table. Always
Y
use the minimum appropriate laser power and gas settings required for a given procedure.
See the Fiber Procedural Setting guide (DOC-GEN-3)or the OR Staff Training Binder (DOC-
FELS-27) for recommended power and gas settings
Note: The CO
laser beam diverges as it exits the distal end of the fiber. Refer to Specification Table
2
for spot size at the fiber tip and beam divergence.
6. Turn off the laser system and gas supply after procedure completion.
7. Disconnect the fiber connector from the laser adapter and remove the fiber from the surgical
instrument.
8. Visually inspect the fiber and verify that no portion is missing.
9. Treat and dispose of used fibers as biohazardous material according to facility guidelines.