by the manufacturer.
may vary from specification. Electrodes
with smaller surface area may cause
tissue irritation.
TO KEEP YOUR DEVICE IN GOOD
WORKING ORDER, OBSERVE THE
FOLLOWING
CAUTIONS:
Caution: Do not immerse your
device in water or place it close
to excessive heat such as a fireplace or
radiant heater or sources of high
humidity such as a nebulizer or kettle as
this may cause it to cease to operate
correctly.
Caution: Keep the device
away from sunlight, as long-
term exposure to sunlight may affect
the rubber causing it to become less
elastic and crack.
Caution: Keep the device
away from lint and dust, as
long-term exposure to lint or dust may
affect the sockets or cause the battery
connector to develop a bad contact.
Caution:
Temperature
Relative Humidity of storage: -
20°C–+40°C,
8%--70%
Temperature & Relative Humidity of
transportation: -20°C–+40°C, 8%--70%
R.H.
Caution: Do not attempt to
open or modify the unit. This
may affect the safe operation of the unit
and will invalidate the warranty.
Note: You may safely use the
stimulator during menstruation,
although it may be a little less
comfortable.
Performance
8. INFORMATION ABOUT
Each
combination of Frequency and Pulse
Width settings which allow for different
ADDITIONAL
sensations
optional electrode pads and help
treating
incontinence.
• Frequency (measured in Hz -
Low frequencies (1-10 Hz) have a
purifying and relaxing effect through
individual contractions.
Medium frequencies (20-50 Hz) can put
a high level of strain on the muscle, thus
promoting the muscular structure.
• Pulse Width (measured in μs -
The Elise unit has pulse widths of 200
to 300 μs. Generally speaking, the
&
higher the pulse width, the more
"aggressive" the stimulation feels, if the
R.H.
pulse width is set high enough, it will
usually elicit a muscle contraction,
which is required for an effective toning
of the pelvic floor muscles.
12
THE
PROGRAMME
SETTINGS
programme
has
through
the
the
different
pulses per second)
millionths of a second)
its
own
probe
or
types
of