General Specification - TensCare Elise Instrucciones De Uso

Ocultar thumbs Ver también para Elise:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

20. GENERAL SPECIFICATION

Waveform
Amplitude
(over 500 Ohm load)
Max intensity
Max pulse energy
Output plug
Channels
Battery
Battery life
Weight
Dimensions
Safety Classification
Environmental Specifications:
Operating:
Storage:
TYPE BF
EQUIPMENT
IP22
Applied Part
Contact duration: At least 10 minutes.
Note: The electrical specifications are nominal and subject to variation from
the listed values due to normal production tolerances of at least 5%.
Asymmetrical rectangular
99.0 mA
+/- 10%
50V zero to peak. Setting 0-99.5 mA in steps of 00.5 mA
Constant voltage over 470-2000 Ohm
Constant current over 160-470 Ohm
OC cutout below 160 Ohm.
Total output limited to 25 μC per pulse
Fully shielded: touch proof mini USB
Single channel
BL-4B Li-ion battery 3.7V
At least 15 hours at 50 mA, 300 μs, 50Hz
90 gms without battery
100 x 65 x 18 mm
Internal power source. Designed for continuous use.
Temperature range: 10 to 35⁰C
Humidity: 20 to 93% RH non-condensing
Atmospheric pressure: 700hPa to 1060hPa
Temperature range: -20 to 70⁰C
Humidity: Up to 93% RH non-condensing
Atmospheric pressure: 700hPa to 1060hPa
Equipment providing a degree of protection against
electric shock, with isolated applied part.
This symbol on the unit means "Refer to Instructions for
Use"
The unit is not water resistant, and should be protected
from liquids.
Complies with EU WEEE regulations
Vaginal and anal electrodes. Optional skin surface
electrode pads. See section 17.
28

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido