clothing, bed linen or in your hair and can only be removed by
thorough ventilation.
Caution!
•
An oxygen supply is only permitted if the WM 24042 O
connection valve is used. Valves of third-party manufacture may
only be used following authorization by the manufacturer,
Weinmann. If unauthorized valves are used, there is a risk of fire.
•
On the topic of oxygen, it is essential to follow the safety instruc-
tions in the instructions for use of your oxygen supply system and
the 0
2
•
Oxygen must be supplied via the breathing mask. Supply via any
other point is not permitted. The quantity of oxygen supplied
must not exceed 4 l/min. If excessive quantities are supplied, there
is a risk of fire.
2.2 Contraindications
CPAP 20e should be used with particular caution or not at all with some diseases.
In individual cases. It is up to the doctor supervising treatment to decide whether
use is indicated. In such cases, use needs to be tightly controlled and the risk
weighed up against the benefits.
This implies:
•
Acute cardiac decompensation
•
Severe cardiac arrhythmias, atrial fibrillation with reduced filling
of the right ventricle.
•
Insufficiency of the right heart or other pulmonary hypertension
•
Severe hypotension, especially in combination with intravascular
volume depletion
•
High risk of barotraumas
•
Bullous lung emphysema
•
Severe deyhdration
oxygen valve.
2
Safety Instructions
EN
13