Switching off the humidifier
There are two ways of switching off the humidifier:
•
Press the humidifier button
midifier symbol
The therapy unit remains switched on.
•
Press the on/off switch on your therapy unit. The humidifier and
the therapy unit are both switched off.
When you switch the therapy unit on again, the humidifier is also
switched on automatically and runs at the heating level of the
previous therapy session.
Automatic switch-off at low levels
If the therapy unit detects that there is not enough liquid in the humidifier, it
switches the humidifier off within 15 minutes. The humidifier symbol
display goes out.
4.3 Softstart setting
The softstart time can be selected in 5-minute increments up to a maximum of 30
minutes. Proceed as follows:
1. Switch on CPAP 20e and press the softstart button
until the display for the softstart time flashes.
2. Now select the softstart time (5-30 minutes in 5-minute increments) using the
on/off switch (+) or the softstart button (-).
If a button is not pressed for 4 seconds, the displayed softstart time will be
saved and the display stops flashing.
The softstart settings are saved when the unit is switched off. Whenever
CPAP 20e is started up, it automatically starts in softstart mode if this was
switched on.
Softstart can be switched on and off at any time by pressing the softstart button
.
EN
22
Operation
of the sleep apnea unit. The hu-
in the display of the sleep apnea unit goes out.
in the
. Hold the button down