Oxygen supply
An oxygen concentrator (e.g. Oxymat), liquid oxygen or an oxygen cylinder with
a suitable pressure reducer may be used. This application must be prescribed by
the patient's doctor.
To prevent oxygen entering the therapy unit, the CPAP 20e automatically gener-
ates a therapy pressure of at least 5 hPa when an oxygen connection valve is at-
tached. You must always follow the safety instructions for handling oxygen as
well as the operating instructions for the oxygen valve and the oxygen unit used.
For safety reasons (risk of fire) the direct introduction
of oxygen into the breathing hose or the mask with-
out any special protective device is not permitted.
However, the supply of up to 4 l/min oxygen to the
breathing mask is possible when using the
WM 24042 oxygen connection valve.