Functional Check; Intervals; Function Check Of Cpap 20E - Weinmann CPAP 20e Descripción Del Aparato E Instrucciones De Manejo

Ocultar thumbs Ver también para CPAP 20e:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Functional check

6.1 Intervals

Carry out a functional check on the CPAP 20e and the CLICK 2 at least every
6 months.
If you find a defect during the functional check, do not use the unit until the
defect has been remedied.

6.2 Function check of CPAP 20e

1. Assemble CPAP 20e with the breathing hose, exhalation system, mask and
power supply cable ready for operation.
2. Push the on/off switch to switch the device on. If Softstart is switched on,
switch it off with the softstart button. The fan conveys air through the breath-
ing hose to the mask and the current pressure in hPa is shown in the display.
3. Close the opening of the mask, e.g. using your thumb or hand.
Compare the pressure shown on the display with the CPAP pressure pre-
scribed. There should only be a divergence of max. ±1.0 hPa.
4. Press the softstart button to switch on softstart. The softstart display appears
and the display shows the softstart time.
The pressure increases slowly and reaches the CPAP pressure after the time
selected. The CPAP pressure is then shown on the display.
5. Switch off the unit using the on/off switch.
If the values or functions are not satisfactory, please send in the unit for repair to
the dealer or the manufacturer.
Functional check
EN
33

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Click 2

Tabla de contenido