5.4 Descaling CLICK 2

If the lower part of the housing and the heater rod are covered in scale:
1. Pour about 150 ml of normal vinegar (5% solution without additives) into the
lower part of the humidifier and leave to stand for one hour.
2. Pour the vinegar solution away.
3. Wipe the housing parts with a clean cloth and rinse thoroughly several times
with water.
4. Then carefully wipe all parts.

5.5 Disinfection CPAP 20e

If necessary, for example after an infectious illness or if the unit has become unusu-
ally contaminated, you can also disinfect the housing, the power supply cable, the
breathing hose, and the humidifier. Please observe the operating instructions for the
disinfectant used. We recommend that you wear suitable gloves when disinfecting
the equipment (e.g. household or disposable gloves).
Unit
Warning!
Electric shock! Before cleaning, remove the power supply cable
from the supply connection of the therapy unit.
Make sure no liquid gets into the housing. Never immerse the
unit in disinfectant or any other liquid. This can damage the unit
and consequently endanger users and patients.
Caution!
Make sure no liquid gets into the pressure glands. This may cause
the unit to malfunction.
The housing, power supply and power supply cable of the CPAP 20e are cleaned
by being simply wiped over with disinfectant. We recommend terralin
EN
30
Hygienic preparation
®
protect.